    題名: 利用 GATE 蒙地卡羅模擬建構雙面板在線式正子造影系統之移變空間響應模型與影像重建;利用 GATE 蒙地卡羅模擬建構雙面板在線式正子造影系統之移變空間響應模型與影像重建
    作者: 馮棋愷;FENG, CHI-KAI
    貢獻者: 光電科學與工程學系
    關鍵詞: GATE模擬;正子斷層造影;列表模式最大期望值演算法;空間變異性;GATE Monte Carlo simulation;PET;LM-MLEM;Spatial variance
    日期: 2026-08-19
    上傳時間: 2026-10-20 15:34:25 (UTC+8)
    出版者: 國立中央大學
    摘要: 質子放射治療在精準醫療中具備極大潛力,但射程不確定性是其臨床
    應用的主要挑戰。雙面板在線式正子造影系統(Dual-Panel On-line PET
    System)能透過原位即時監測質子誘發之短半衰期同位素(如 15O),提供可靠
    的射程驗證。然而,受限於系統非連續型偵測器的幾何排列,其投影數據嚴
    重欠缺,導致三維影像重建易產生嚴重的有限角度假影與離焦模糊,降低了
    定性與定量的準確性。
    為改善此幾何限制造成的影像劣化,本研究利用 GATE 蒙地卡羅模擬
    軟體,建構與臨床相符的系統模型。研究中首先進行取樣完整性與系統靈敏
    度的量化分析, 結果顯示,取樣完整性與靈敏度會隨偏離視野中心及加大探
    頭間距而急遽下降。
    接著,透過三維空間的格點掃描實驗,獲取不同空間位置的點射源投影
    數據,藉此建構具備空間變異性的影像系統矩陣。
    最後,本研究結合列表模式期望值最大化演算法(List-Mode Maximum
    Likelihood Expectation Maximization, LM-MLEM)與移變理查森-露西反卷積
    演算法(Shift-Variant Richardson-Lucy Deconvolution Algorithm, SV-RL)進行
    影像重建與比較,以評估解析度恢復的效果;Proton radiotherapy holds great potential in precision medicine, but range
    uncertainty remains a major challenge in its clinical application. The Dual-Panel
    On-line PET System can provide reliable range verification through in-situ, realtime monitoring of proton-induced short-half-life isotopes (such as 15O). However,
    restricted by the geometric arrangement of the system′s non-continuous detectors,
    its projection data is severely deficient. This deficiency makes three-dimensional
    image reconstruction highly susceptible to severe finite-angle artifacts and
    defocus blur, thereby degrading qualitative and quantitative accuracy.
    To address the image degradation caused by these geometric limitations, this
    study utilized the GATE Monte Carlo simulation software to construct a system
    model consistent with clinical settings. Initially, a quantitative analysis of
    sampling completeness and system sensitivity was performed. The results
    indicated that both sampling completeness and sensitivity decline sharply as the
    position deviates from the center of the field of view (FOV) and as the distance
    between the detector panels increases.
    Subsequently, a three-dimensional grid scan experiment was conducted to
    acquire projection data from point sources at various spatial locations, thereby
    constructing an imaging system matrix with spatial variation.
    Finally, this study combined the List-Mode Maximum Likelihood
    Expectation Maximization algorithm and the Shift-Variant Richardson-Lucy
    Deconvolution Algorithm to perform image reconstruction and comparison,
    aiming to evaluate the effectiveness of resolution recovery.