中大機構典藏-NCU Institutional Repository-提供博碩士論文、考古題、期刊論文、研究計畫等下載:Item 987654321/53054
English  |  正體中文  |  简体中文  |  全文笔数/总笔数 : 80990/80990 (100%)
造访人次 : 42650101      在线人数 : 968
RC Version 7.0 © Powered By DSPACE, MIT. Enhanced by NTU Library IR team.
搜寻范围 查询小技巧:
  • 您可在西文检索词汇前后加上"双引号",以获取较精准的检索结果
  • 若欲以作者姓名搜寻,建议至进阶搜寻限定作者字段,可获得较完整数据
  • 进阶搜寻


    jsp.display-item.identifier=請使用永久網址來引用或連結此文件: http://ir.lib.ncu.edu.tw/handle/987654321/53054


    题名: 結構振型探討牙科植體術後臨床骨缺損;The Mode shape of structures Assess Clinical Bone Defects after Dental Implantation
    作者: 謝易珊;Yi-san Hsieh
    贡献者: 機械工程研究所
    关键词: 牙科植體;骨缺損;共振頻率;模態分析;簡諧分析;局部模態振型;Dental implant;Bone defect;Resonant frequency;Modal analysis;Harmonic analysis;Local mode shape
    日期: 2012-01-11
    上传时间: 2012-06-15 19:38:21 (UTC+8)
    摘要: 本研究以結構振型探討牙科植體術後臨床骨缺損檢測課題,透過數值分析方法,找出植體/下顎骨局部模態植體振型及共振頻率,以利於評估骨缺損程度。研究分兩部份,第一部份為有限元素法分析,使用結合與摩擦兩種植體與下顎骨接觸條件,分別模擬臨床之植牙骨整合與以人造下顎骨進行實驗之情況,由模態分析獲得植體/下顎骨之自然頻率與振型,透過簡諧分析分別得到植體在頰舌側及近遠心側之響應。第二部份為聲能激振-位移響應實驗,以喇叭之聲能激振植體/下顎骨結構,並以非接觸式電容式位移計量測振動響應訊號,後續和Osstell III檢測儀量測結果作比較。 模擬結果顯示,結構產生植體局部模態振型時,檢測骨整合穩固度成效較明顯,而非接觸式聲能激振-位移響應實驗,受限於高頻響應位移小,以及位移計量測動態範圍、訊雜比低,對於高頻局部模態實驗效果不佳,僅能測得下顎骨低頻整體共振,不同缺損型態之共振頻率差異不大;另以Osstell III檢測儀可明確評估骨缺損程度。故上述模擬與實驗結果可證實需以植體局部模態之共振頻率進行骨缺損檢測。The aim of this study is to assess clinical bone defects between an implant and jaw bone after dental implantation by examining the mode shape of structures. Different severity of bone defects was evaluated through structure resonant frequencies and corresponding mode shapes of the implant and jaw bone by using numerical analysis and experimentation. This study consists of two parts. First, the assumption of two kinds of boundary conditions, bonding and rubbing, was applied to simulate osseointegration in the clinical dentistry and the in-vitro bone defect model, respectively, in finite element analysis. Natural frequencies and their mode shapes of the implant/jaw were computed by the modal analysis. During the harmonic analysis, the response displacement versus frequency of implant in the buccolingual and mesiodistal directions were defined. Secondly, the structural resonant frequencies were measured by the acoustic excitation-displacement response procedure, and then this result was compared with using the detection of Osstell mentor. The simulation results show that the structure local mode corresponding high-frequency resonance can be used to examining bone imperfection remarkably. Limited by extremely tiny response displacement, measuring dynamic range of the capacitive displacement sensor, the acoustic excitation-displacement response measurement can only acquire the structure global mode of the mandible corresponding to low-frequency resonance. In addition, Osstell mentor can assess bone defects effectively. Therefore, the above-mentioned simulations and experimental results prove that the local mode is promising to evaluate the defect severity for the osseointegration of dental implantation.
    显示于类别:[機械工程研究所] 博碩士論文

    文件中的档案:

    档案 描述 大小格式浏览次数
    index.html0KbHTML585检视/开启


    在NCUIR中所有的数据项都受到原著作权保护.

    社群 sharing

    ::: Copyright National Central University. | 國立中央大學圖書館版權所有 | 收藏本站 | 設為首頁 | 最佳瀏覽畫面: 1024*768 | 建站日期:8-24-2009 :::
    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 隱私權政策聲明