English  |  正體中文  |  简体中文  |  全文筆數/總筆數 : 80990/80990 (100%)
造訪人次 : 42687695      線上人數 : 1455
RC Version 7.0 © Powered By DSPACE, MIT. Enhanced by NTU Library IR team.
搜尋範圍 查詢小技巧:
  • 您可在西文檢索詞彙前後加上"雙引號",以獲取較精準的檢索結果
  • 若欲以作者姓名搜尋,建議至進階搜尋限定作者欄位,可獲得較完整資料
  • 進階搜尋


    請使用永久網址來引用或連結此文件: http://ir.lib.ncu.edu.tw/handle/987654321/93530


    題名: 穿戴式低能量脈衝電磁場設備開發及其應用;Development and application of a wearable low-energy pulsed electromagnetic field device
    作者: 張博翔;Chang, Po-Hsiang
    貢獻者: 電機工程學系
    關鍵詞: 脈衝電磁場;自主神經系統;心率變異度;可穿戴式裝置;Pulsed electromagnetic field;Autonomic nervous system;Heart rate variability;Wearable device
    日期: 2023-11-01
    上傳時間: 2024-09-19 17:11:24 (UTC+8)
    出版者: 國立中央大學
    摘要: 自主神經系統在人體健康中扮演著重要的調節作用。自主神經系統(Autonomic nervous system,ANS)失調可能導致許多疾病。近年來,運用脈衝電磁場(Pulsed electro magnetic fields,PEMF)調節自主神經系統的平衡引起了極大的關注。PEMF治療不需要直接接觸用戶的身體,是一種非侵入性、非接觸和非熱療法,在長時間使用下不會對皮膚造成損傷。然而,目前的PEMF設備由於需要大尺寸的磁場線圈來產生足夠的磁場強度,因此設備體積較為龐大。本研究提出了一種使用軟性印刷電路板(Flexible printed circuit board,FPCB)製造的多層PEMF裝置。所提出系統的刺激線圈具有可彎曲、薄、輕、磁場強度高的優點。PEMF系統已應用於通過姿勢誘導的交感/副交感變化調節ANS平衡的傾斜桌測試中。PEMF刺激顯示在調節ANS平衡方面具有顯著的改善作用。與非PEMF條件相比,頻譜心率變異指標LF、HF和LF / HF比值分別為62.02 ± 19.35 vs. 71.17 ± 14.34、37.98 ± 19.35 vs. 28.83 ± 14.34和2.62 ±2.41 vs. 4.03 ± 3.96,表明PEMF刺激對ANS調節具有效果。所提出的可穿戴式PEMF設備具有良好的穿戴性,並可以與衣服結合使用,這種技術能夠在未來的人體研究中實現長期刺激。;The autonomic nervous system (ANS) plays a crucial role in maintaining human health, and dysautonomia can lead to various diseases. Pulsed electromagnetic field (PEMF) therapy has emerged as a promising approach for regulating ANS balance. However, existing PEMF devices are typically bulky due to the large size of the magnetic coil required to generate sufficient magnetic strength for stimulation. In this study, we propose a multi-layer PEMF device fabricated using flexible printed circuit board (FPCB) technology. The stimulation coil of our device offers advantages such as flexibility, thinness, lightweight, and high magnetic field intensity.
    This study applied our PEMF system to regulate ANS balance by conducting tilt-table tests to induce posture-induced sympathetic/parasympathetic changes. The results demonstrated a significant improvement in ANS balance with PEMF stimulation. When comparing the PEMF and non-PEMF conditions, the spectral heart rate variability (HRV) indices, including LF (low frequency), HF (high frequency), and LF/HF ratio, showed significant changes. Specifically, LF values were 62.02 ± 19.35 vs. 71.17 ± 14.34, HF values were 37.98 ± 19.35 vs. 28.83 ± 14.34, and LF/HF ratio values were 2.62 ± 2.41 vs. 4.03 ± 3.96, respectively. These results highlight the effectiveness of PEMF stimulation in regulating ANS activity.
    The proposed wearable PEMF device offers excellent wearability and can be seamlessly integrated with clothing, which enables the long-term stimulation in future human studies.
    顯示於類別:[電機工程研究所] 博碩士論文

    文件中的檔案:

    檔案 描述 大小格式瀏覽次數
    index.html0KbHTML32檢視/開啟


    在NCUIR中所有的資料項目都受到原著作權保護.

    社群 sharing

    ::: Copyright National Central University. | 國立中央大學圖書館版權所有 | 收藏本站 | 設為首頁 | 最佳瀏覽畫面: 1024*768 | 建站日期:8-24-2009 :::
    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 隱私權政策聲明